Melting range determination of semi-crystalline polymers according to EN ISO 3146

BÜCHI Labortechnik AG

According to the official regulation EN ISO 3146, polymers do usually not have a sharp melting point as known for low-molecular substances. Here, we demonstrate the feasibility to employ the M-565 for the melting range determination of semi-crystalline polymers by capillary tube method according to EN ISO 3146.
Semi-crystalline poly(tetrafluoroethylene) (PTFE was chosen as test compound for the melting range measurements. For the analysed PTFE samples, we report a narrow melting range between 322.9 ± °C and 326.0 ± 0.1 °C.
With the M-565 the melting process can be measured and recorded without modification of the instrument

Laadi alla rakendus Laadi alla rakendus Rakenduse ülevaade Võta meiega ühendust

Application Download Form

* Need väljad on kohustuslikud.

** Võtke arvesse, et teie isikuandmeid säilitatakse turvaliselt ja kooskõlas isikuandmete kaitse üldmäärusega (GDPR) (EL) 2016/679 füüsiliste isikute kaitse kohta isikuandmete töötlemisel ja selliste andmete vaba liikumise kohta. Teil on õigus nõuda meilt juurdepääsu kõigile teie salvestatud isikuandmetele, nende parandamist, isikuandmete kustutamist või töötlemise piiramist, samuti saate oma nõusoleku igal ajal tagasi võtta. Isikuandmeid ei jagata väljaspool meie kontserni ega müüda teistele ettevõtetele. Isikuandmeid võidakse turundus- ja kommunikatsioonieesmärkidel kasutada kuni nõusoleku tühistamiseni.